Mediastinum      (MRI CHEST, w/wo)                                       (45m – 12/04)
PulseSeq
TR
TE
Flip
BW
Matrix
Slice
SAT
NEX
FOV
Other
1. Cor SSFSE
na
80
na
62
256x160
8 x 0
None
.5
36x36
 
2. Ax SSFSE
na
80
na
62
256x128
8 x 0
None
.5
32x24
 
3. Axial Blood Suppressed
R-R
20
na
62
256x128
8 x .5
None
1
26x26
BrHold
4. Axial T2 Blood Suppressed
R-R
80
na
62
256x128
8 x .5
Fat
1
26x26
BrHold
#5. Axial FSPGR timing
min
min
60°
12
256x128
20
S, I
1
32x24
At Pulm Bifurc; 5 cc
*6. Ax LAVA
min
min
15°
31
256x128
4
SpecFat
.5
26x26
Pre/post-Gd x2; SliceZip
7. Axial FastCine
min
min full
30°
31
256x128
8 x 1
None
.75
28x28
BrHold; 12 views/seg

Ax FIESTA FS

20

min

70°

15

192X288

5 x 0

Fat

1

32x24

Breathhold

EKG Gating and Respiratory Bellows Must be Applied
#Time to descending Aorta. 60 slices at same location. .
*30cc Gad.


Copyright © Thomas Jefferson University. All Rights Reserved.
The Thomas Jefferson University web site, its contents and programs, is provided for informational and educational purposes only and is not intended as medical advice nor is it intended to create any physician-patient relationship. Please remember that this information should not substitute for a visit or a consultation with a health care provider. The views or opinions expressed in the resources provided do not necessarily reflect those of Thomas Jefferson University, Thomas Jefferson University Hospital, or the Jefferson Health System or staff.